CyanoFluor Handheld HAB Indicator FAQs
Print

What can I do if my Calibration Check Standard fails?

Wipe the windows in the sample compartment with a clean dry end of a Q-tip to lightly wipe away any debris or residue from moisture that may be interfering with the measurement. 
Rerun the Calibration Check Standard 5 times while on a flat surface.  
     If the Calibration Check passes 3 out of 5 times, the instrument is ready for use.
     If the Calibration Check fails 3 out of 5 times, repeat the cleaning procedure
Rerun the Calibration Check Standard 5 times while on a flat surface.  
     If the Calibration Check passes 3 out of 5 times, the instrument is ready for use.
     If the Calibration Check fails 3 out of 5 times, contact Turner Designs