C-sense pCO2 Sensors (FAQ)
Print

Does the C-sense have to be corrected for hydrostatic pressure?

No. Hydrostatic pressure has no effect.

Does the C-sense need to be corrected for salinity?

No.  Salinity has no effect on the C-sense measurement as it is an aqueous measurement.

Does the C-sense need corrections for temperature and total dissolved gas pressure (TDGP)?

The C-sense does not need to be corrected for temperature, it is compensated for in the sensor. However for increased accuracy, it should be corrected for TDGP.   Each C-sense is calibrated at 20 degrees C and an atmospheric pressure at or near 101.3kPa (pressure at the time of calibration is stated on the calibration sheet which ships with the C-sense).   Corrections can be made, assuming TDGP is known, using the equation below.

    pCO2 actual = (pCO2 measured x TDGP) / calibration pressure

Under what conditions should a user measure TDGP?

A user needs to measure TDGP when the atmospheric pressure is significantly different than the atmospheric pressure at the time of calibration.  This situation arises during extreme weather events and when sampling at elevation.  The largest pressure swings caused by weather are on the order of 3% deviation from standard sea level pressure.  Independent of atmospheric pressure, TDGP in the ocean is primarily controlled by the amount of dissolved oxygen.  Therefore, if oxygen levels at your intended sampling site are anomalous, you are advised to measure TDGP.

What kind of error can be expected if TDGP is not measured?

For every 1% change in TDGP from the pressure during calibration, there results an ~1% error in CO2.  To put this into context, the accuracy of the C-sense is 3% of full scale, so depending on the application, you may or may not need to correct for TDGP.

How does the C-sense measure at the specific wavelength absorbed by CO2?

As a proxy to CO2 concentration, absorption of a light source passing through the sample gaseous headspace is measured.  In order to limit the wavelengths emitted by the light source to those specific to CO2 absorption, a filter is used.

Does the C-sense sample wet air or dry air?

Because the C-sense does not measure, or compensate for water vapor entering the headspace, it measures a wet sample.

What is the warmup time for the instrument?

The warm-up time, specific to the operation of the detector, is between 45 seconds and 3 minutes.  If the instrument has been powered down and allowed to cool, the user should wait the full 3 minutes.  If the instrument is sampling frequently, it won’t cool as much between samples, and therefore, the warm up time is less.  When choosing a warm-up time, the ambient temperature of the water being measured should be considered.  In order to prevent recording measurements before the detector has warmed, the instrument will not output data for 45 seconds after power is applied.

What it the equilibration time for the instrument?

Equilibration times for pCO2 sensors are often stated as a t63 level.  This is defined as the time it takes for the instrument to reach 63% of equilibration.  For the C-sense, this value is ~4 minutes.
The equilibration time can be reduced with the use of the Water Pumped Head (P.N. 2400-700).

Can the C-sense be integrated with a SeaBird Electronics CTD?

The C-sense outputs a 0-5 VDC signal proportional to the pCO2 concentration.  There are multiple Sea-Bird Electronics CTD’s that can record this signal, convert it to a concentration in ppm, and write it to a file.  Information on how to integrate the C-sense to a SeaBird Electronics SBE19 CTD is available here or under the “Documentation” tab on the C-sense webpage.

How often does the C-sense need to be calibrated? If the calibration is a year old, but the instrument has never been deployed, does the instrument need to be recalibrated?

With regular use, the C-sense needs to be calibrated once a year.  If the instrument has never been deployed, or deployed infrequently followed by the recommended cleaning techniques, there is no need to re-calibrate.

Is bio-fouling an issue for the C-sense and if so, how can it be mitigated?

Bio-fouling is an issue for most submersible instrumentation measuring natural environments, both fresh and saline.  In the case of the C-sense, accumulation of organisms on the membrane has two potential but competing effects.  The presence of the accumulation can decrease the equilibration time by impeding the transfer of gasses.  Alternatively, the organisms can emit CO2 as a result of respiration, artificially elevating the CO2 concentration.  Turner Designs offers accessories for the C-sense that help reduce the effects of bio-fouling.  Copper Cutouts (P/N 2400-506) can be applied to the sensor head outside the membrane.  This prevents organisms from growing on the instrument head.  The Water Pumped Head (P/N 2400-700) accessory will inhibit bio-fouling in two ways, first by significantly reducing the amount of available light for photosynthesis, secondly by creating shear stress that prevents organisms from settling on the membrane.  The Water Pumped head has the additional advantage of reducing equilibration times.  The two accessories should not be used together so it is recommended that the user decide on the necessity of the Water Pumped head first, and plan accordingly.

What material are the screws in the C-sense made of?

The screws are 3-16 stainless steel.