Ballast-Check 2 FAQs
Print

How can I see the logged data on the Ballast-Check 2?

Data can be viewed on the Ballast-Check 2’s display or downloaded to a computer for viewing.  To view data on the Ballast-Check 2’s display from the HOME screen, use the left/right arrow keys to scroll through sample numbers until you reach the sample number you wish to view.  Then use the up/down arrow keys to scroll through logged data for that sample number.  You can view the date, time, risk, ship/tank values recorded and the Abundance and Activity value saved for that sample.  You can also view these parameters using the up/down arrow keys right after a measurement has completed.

What can I do if my Calibration Check Standard fails?

Wipe the windows in the sample compartment with a clean dry end of a Q-tip to lightly wipe away any debris or residue from moisture that may be interfering with the measurement. 
Rerun the Calibration Check Standard 5 times while on a flat surface.  
     If the Calibration Check passes 3 out of 5 times, the instrument is ready for use.
     If the Calibration Check fails 3 out of 5 times, repeat the cleaning procedure
Rerun the Calibration Check Standard 5 times while on a flat surface.  
     If the Calibration Check passes 3 out of 5 times, the instrument is ready for use.
     If the Calibration Check fails 3 out of 5 times, contact Turner Designs